Standard: 790 mm x 815 mm x 300 mm
SmartScope® Quest™ 800 is an extremely accurate floor model coordinate measurement system incorporating scanning probe, touch probe, micro-probes, and the patented TeleStar® Plus TTL laser for versatility and accuracy.
The Quest 800, with its extensive 790 x 815 x 300 mm XYZ measurement volume (and available 400 mm Z-travel), is designed to measure very large parts. A solid granite bridge platform rests on a sturdy steel support structure and the bridge design limits measured part motion to the Y axis, offering measurement stability. Quest 800 offers:
Accurate video metrology – TeleStar® telecentric 10:1 zoom optics for the highest level of optical performance
Superb illumination for the best video measurements – Standard profile light, coaxial surface light, and SmartRing™ light illuminate parts from all angles multisensor dimensional measurement
Multisensor versatility – Optional touch probe, off-axis DRS™ laser, on-axis TeleStar TTL interferometric laser, micro-probes, SP25 continuous contact scanning probe, PH10 motorized probe head, and 4th and 5th axis rotary indexers
790 mm x 815 mm x 300 mm
Extended Z axis, 400 mm
Fully telecentric motorized zoom with AccuCentric®
auto-compensation, 1x standard lens
Focus Grid Projector: LED source
Laser Adapter: Allows for field retrofit of TTL Laser (includes laser pointer)
Replacement / Laser Lenses: 0.45x, 0.5x, 2x, 4x
Substage LED profile, coaxial LED surface,
SmartRing™ LED ring light
Flexible SmartRing light for 0.45x and 0.5x lenses, White Fiber-Optic Ring (LED or Tungsten) mounted
below or in lieu of SmartRing (1x, 2x, and 4x only)
| Standard | Optional | |
|---|---|---|
| XYZ Travel | 790 x 815 x 300 mm | Extended Z-axis: 400 mm |
| XYZ Scale Resolution | 0.1 μm | 0.05 µm |
| Drive System | DC servo with 4-axis control (X, Y, Z, zoom) and multifunction handheld controller | XY liquid cooled linear motor drives |
| Worktable | Hardcoat anodized with fixture holes and removable stage glass, 75 kg recommended max payload | |
| Rotary Axis | Miniature Servo Rotary (MSR™), MicroTheta Rotary (MTR™), Heavy Duty Rotary (HDR) | |
| Optics1 | Fully telecentric motorized zoom with AccuCentric® auto-compensation, 1x standard lens | Focus Grid Projector: LED source Laser Adapter: Allows for field retrofit of TTL Laser (includes laser pointer) Replacement / Laser Lenses: 0.45x, 0.5x, 2.0x, 4.0x |
| Illumination | Substage LED profile, coaxial LED surface, SmartRing™ LED ring light | Flexible SmartRing light for 0.45x and 0.5x lenses, White Fiber-Optic Ring (LED or Tungsten) mounted below or in lieu of SmartRing (1x, 2x, and 4x only) |
| Metrology Camera | 2 megapixel monochrome digital metrology camera | |
| Field of View | 8.0 mm x 6.0 mm | 0.45x: 16.0 x 12.0 mm |
| Minimum Feature Size2 | 2.5 µm | 0.45x: 5.0 µm |
| Working Distance | 68 mm | 0.45x: 200 mm 0.5x: 130 mm 2x: 34 mm 4x: 19 mm |
| Sensor Options3 | Tactile: TP20 or TP200 Touch Probe, SP25 Scanning Probe, Feather Probe™, PH10 articulating probe head4 Non-Contact: TeleStar Plus Interferometric TTL Laser, DRS™ Laser, Rainbow Probe™ |
|
| Software | ZONE3 Express metrology software, OGP® Portal |
Metrology Software: ZONE3 Prime or Pro Productivity software: OGP EVOLVE® Software (Design, Manufacturing, SmartProfile®, SPC) Offline software: ZONE3 |
| System Controller | Windows® based with up-to-date processor and on board networking/communication ports | High-Performance Controller, Sub-Controller Network Isolator |
| Controller Options | 24” flat panel LCD monitor or dual 24” flat panel LCD monitors, keyboard, 3-button mouse (or user supplied) | |
| Accessories | Digital I/O Automation Interface | |
| Power Requirements | 100-200 VAC or 200-240 VAC, 50/60 Hz, 1 phase, 1000 W | |
| Safe Operating Environment | 15-30 °C, non-condensing | |
| Rated Environment | Temperature 18-22 °C, stable to ± 1 °C; max rate of change 1 °C / hour; max vertical gradient of 1 °C / meter; 30-80% humidity; vibration <0.001g below 15 Hz | |
| XYZ Volumetric Accuracy | E3 = (2.8 + 5L/1000) µm | |
| XY Area Accuracy | E2 = (2.0 + 5L/1000) µm | E2 = (1.4 + 5L/1000) µm (requires optional 0.05 µm scales) |
| Z Linear Accuracy | E1 = (2.5 + 5L/1000) µm | E1 = (1.5 + 5L/1000) µm (requires optional TeleStar Plus TTL Laser, DRS Laser, or Touch Probe) |
Last updated: October 2026
OGP Authorized Representatives are metrology specialists, qualified to evaluate your measurement requirements and recommend the solution that best fits your needs.
Your local representative is ready to help you with product details, demonstrations, training, or purchase information.